Page 96

proefschrift_Schols_SLV

Chapter 7                              Figure 7.2  Optical fiber probe measurement + detail of fiber probe tip    Data collection  Before specimen measurements, a  reflectance baseline was recorded with  spectralon  (Optical‐grade  spectralon  reference  material;  Labsphere,  Inc.,  North  Sutton,  New  Hampshire USA). Spectralon is used very commonly as reference spectrum for baseline  correction as it behaves as a nearly perfect cosine‐diffused reflector from 250‐2500 nm.  Baseline  values  were  measured  with  the  fiber  probe  in  direct  contact  with  the  spectralon  surface.  During  the  baseline  recordings  the  light  source  was  set  at  50%  power  capacity. The same  fiber probe  that was used  for  the baseline  recordings was  also used for the ex vivo tissue measurements. Overall tissue reflection was much lower  than  spectralon  reflectance.  Hence,  the  light  source  was  turned  up  to  100%  during  spectroscopic measurements on  tissue, thus  achieving a  better dynamic range during  tissue measurements  (light  source  linearity  was  verified  separately).  The  fiber  probe  was gently brought into direct contact with the tissue surface by hand as demonstrated  in  Figure  7.2.  Between  measurements  of  different  tissue  types  the  fiber  probe  was  cleaned with saline.   Spectra of  all  tissue  types were  acquired  from  fresh human  tissue,  immediately  after  surgical  removal  while  the  specimen  was  still  within  the  operating  theatre.  Several  spectra  were  recorded  for  each  tissue  type:  healthy  colonic  tissue,  mesenteric  fat,  arteries, and veins; if present in the specimen, also: ureter, palpable tumor in lumen of  intestine.  At  least  two  different  locations  were measured  per  tissue  type  and  each  location was measured five times. This provided a total of at least ten spectra per tissue  94 


proefschrift_Schols_SLV
To see the actual publication please follow the link above